Honey, adulteration

The Honey Trap: Analytical Technology Makes Food Fraud Easier to Catch

By Christopher Brodie
1 Comment
Honey, adulteration

Protecting product integrity remains a challenge for honey manufacturers and suppliers.

One of the key benefits of LC-IRMS is the ability to determine carbon isotope fingerprints for each individual sugar component, comparable to EA-IRMS, which provides a carbon isotope fingerprint for all sugars in the honey. In this way, a comparison of δ13C of fructose and glucose, as well as the detection of other unknown sugars and the honey’s sugar pattern, can be determined within a single sample analysis. A representative LC-IRMS chromatogram for honey is shown in Figure 2.

Honey Adulteration, LC-IRMS chromatogram
Figure 2. Representative LC-IRMS chromatogram showing sugar analysis of honey and chromatographic conditions.

Table I reports data from eight honey samples analyzed using LC-IRMS and EA-IRMS to identify adulterated and unadulterated honey. A Thermo Scientific LC IsoLink IRMS System was used to analyze the honey samples in a comparative study with EA-IRMS.6 These data show that two cases of adulteration (Samples 1 and 5) were identified using EA-IRMS and four adulterated honeys (Samples 1, 2, 5, and 8) were detected using LC-IRMS.

Honey samples, analysis
Table 1. δ13C values of eight honey samples analyzed by LC-IRMS and EA-IRMS.

Despite exhibiting a normal fructose/glucose ratio of 1.07, honey sample 1 was suspected to be adulterated due to the less negative δ13C of the honey, relative to the protein. For honey sample 2, the significantly lower glucose content relative to fructose, and its resulting fructose/glucose ratio of 0.65 gives a clear indication of adulteration. Additionally, the 2.7‰ more negative δ13C value of fructose than glucose provides additional evidence to support this conclusion. Honey sample 5 is also adulterated, identified from the high fructose/glucose ratio of 4.53. Honey sample 8 was also identified as adulterated, from the difference of 1.3% in the δ13C values of glucose and fructose, as well as the high fructose/glucose ratio of 2.17. However, this conclusion could not be reached using the EA-IRMS data, as a δ13C difference of 0.3‰ (i.e., 1.5% adulteration with C4 derived sugar) is within the uncertainty of the method according to the AOAC method 998.12. However, the data clearly show that the combination of EA-IRMS and LC-IRMS can detect a wide range of honey adulteration, even at low levels, demonstrating the power of isotope fingerprints in detecting food adulteration.

Summary

EA-IRMS is a robust and proven technology for isotopic ratio measurement and can be used as a routine and reliable approach for the identification of honey adulteration, following the AOAC 988.12 official method. For low concentration adulteration, compound specific analyses using LC-IRMS benefits from enhanced analytical sensitivity and the ability to identify individual sugar isotope fingerprints. With adulteration a significant issue for honey manufacturers and suppliers, carbon isotope fingerprinting provides the industry with a powerful, robust and versatile solution to the challenge of safeguarding product integrity.

Continue to the sidebar on page 3 below: Follow the Honey: Safeguarding Food Integrity Using Integrated Informatics